Serum serial number incorrect9/9/2023 ![]() Intermittent failures may occur when calibrating the T4 assay due to observed errors in one or more replicates. Issue Number 3 - The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not calibrate T4. VB12 dilution results generated with expired T3/T4/VB12 Ancillary Reagent are not reported with a flag. T3 and T4 results generated using expired T3/T4/VB12 Ancillary Reagent are reported with a result flag, "OBS extended". Issue Number 2 - The Atellica IM 1300 Analyzer and IM 1600 Analyzer may use the T3/T4/VB12 Ancillary Reagent after expiration. When an Atellica Magline Transport cover in the front or back of an analyzer is opened, carrier motion may continue and the operator may not be alerted that a cover is open. Issue Number 1 - The cover interlocks on the Atellica Magline Transport for the Atellica CH 930 Analyzer, IM 1300 Analyzer, and IM 1600 Analyzer may not work as expected. ![]() Siemens Healthcare Diagnostics has identified the following issues with the Atellica Solution, which are installed with Atellica Solution software (SW) versions 1.14.2 (SMN 11316393) or lower.